LAVA ULTIMATE CAD-CAM RESTORATIVE FOR CEREC 2914A2-LT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-04-04 for LAVA ULTIMATE CAD-CAM RESTORATIVE FOR CEREC 2914A2-LT manufactured by 3m Espe Dental Products.

Event Text Entries

[71620811] Based on the available records, it is not clear what role, if any, the lava ultimate restoration, played in this event. Other factors, such as prior tooth history and dental treatments, may have played a role in the need for root canal therapy.
Patient Sequence No: 1, Text Type: N, H10


[71620812] This patient record was identified during post-market surveillance activities. Records supplied by the dental office indicate that a male patient ((b)(6)) required root canal therapy on tooth #3. This tooth had received a lava ultimate crown on (b)(6) 2014, because recurrent decay had occurred beneath a prior temporary crown; details on when the temporary crown was placed were not provided. On (b)(6) 2014, the lava ultimate crown debonded and was re-cemented. On (b)(6) 2014, the crown fractured and a new lava ultimate crown was placed. Records indicate that at this time, very little tooth structure remained; the dental professional noted that if the crown debonded again, it was possible that a root canal and core build-up might be necessary. On (b)(6) 2014, the lava ultimate crown debonded and was re-secured with a temporary cement. On (b)(6) 2014, a crack was identified in the tooth and patient was referred to endodontist for root canal (which was likely conducted on (b)(6) 2014). On (b)(6) 2014, a new non-3m crown was placed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005174370-2017-00028
MDR Report Key6456833
Report SourceHEALTH PROFESSIONAL
Date Received2017-04-04
Date of Report2017-03-08
Date Mfgr Received2017-03-08
Date Added to Maude2017-04-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANGIE DRAPER
Manufacturer Street2510 CONWAY AVENUE
Manufacturer CityST. PAUL MN 551441000
Manufacturer CountryUS
Manufacturer Postal551441000
Manufacturer G13M ESPE DENTAL PRODUCTS-IRVINE
Manufacturer Street2111 MCGAW AVENUE
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal Code92614
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-2052-2015
Event Type3
Type of Report3

Device Details

Brand NameLAVA ULTIMATE CAD-CAM RESTORATIVE FOR CEREC
Generic NameMATERIAL, TOOTH SHADE, RESIN
Product CodeEBF
Date Received2017-04-04
Catalog Number2914A2-LT
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M ESPE DENTAL PRODUCTS
Manufacturer Address2510 CONWAY AVENUE ST. PAUL MN 551441000 US 551441000


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-04-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.