SUPPORT ARM 177 61481720

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-06 for SUPPORT ARM 177 61481720 manufactured by Maquet Critical Care Ab.

Event Text Entries

[72194458] No parts were returned. Investigation was based on a picture we had received. According to the picture, the support arm broke at the joint nearest to the bracket. The support arm is a casting and it most probably developed a crack at an earlier occasion either by overloading, or impact, and this led to the reported breaking. The support arm has been successfully tested for mechanical strength and is designed according to standard. It is unknown under which circumstances the support arm broke.
Patient Sequence No: 1, Text Type: N, H10


[72194459] It was reported that the support arm broke at the lower end. There was no patient involvement reported. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010042-2017-00149
MDR Report Key6463301
Date Received2017-04-06
Date of Report2017-04-06
Date of Event2017-03-23
Date Mfgr Received2017-03-30
Date Added to Maude2017-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Manufacturer G1MAGNUS LINDQVIST
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSUPPORT ARM 177
Generic NameSUPPORT, ARM
Product CodeIOY
Date Received2017-04-06
Catalog Number61481720
Lot Number2010/02
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CRITICAL CARE AB
Manufacturer AddressSOLNA SW


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-06

© 2026 FDA.report