IMAGE MANAGEMENT SYSTEM 660HD-E 72204668

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-06 for IMAGE MANAGEMENT SYSTEM 660HD-E 72204668 manufactured by Smith & Nephew, Inc..

Event Text Entries

[72132352] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[72132353] It was reported the screen went black on the image management system 660hd-e. This occurred during the procedure. A backup device was readily available. There were no delays or patient injuries reported.
Patient Sequence No: 1, Text Type: D, B5


[74732502] Device investigation narrative - complaint of blank screen could not be reproduced. Unit passed functional testing during 6 hour burn-in including touchscreen functions and capturing video stills and clips. Display remained constant throughout functional testing and burn-in. All video inputs/outputs normal and unit's software performed as expected. Unit was tested in both pal and ntsc formats and both analog and digital video outputs. No problem found. A review of the device history record for serial number (b)(4) shows that this unit passed all acceptance criteria and was compliant upon release for distribution on february of 2017. No containment or corrective actions are recommended at this time.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1643264-2017-00190
MDR Report Key6464992
Date Received2017-04-06
Date of Report2017-05-04
Date of Event2017-03-16
Date Mfgr Received2017-05-03
Device Manufacturer Date2017-02-21
Date Added to Maude2017-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street7000 WEST WILLIAM CANNON DRIVE
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer G1SMITH & NEPHEW, INC.
Manufacturer Street76 S. MERIDIAN AVE.
Manufacturer CityOKLAHOMA CITY OK 731076512
Manufacturer CountryUS
Manufacturer Postal Code731076512
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMAGE MANAGEMENT SYSTEM 660HD-E
Generic NameDEVICE, DIGITAL IMAGE STORAGE, RADIOLOGICAL
Product CodeLMB
Date Received2017-04-06
Returned To Mfg2017-04-10
Catalog Number72204668
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address76 S. MERIDIAN AVE. OKLAHOMA CITY OK 731076512 US 731076512


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-06