MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-07 for CAPTURE-R READY INDICATOR RED CELLS 0006428 manufactured by Immucor, Inc..
[72510615]
The full number for the product in question is bk020053 (1/27/2003). The immucor laboratory tested retention product on 03apr2017, which performed as expected. Immucor technical support used a remote electronic connection method to assess the instrument test well images in question on 28mar2017, and determined that the instrument test well images in question were visually negative. An immucor field service engineer (fse) visited the customer site to assess the testing instrument on 28mar2017. The fse replaced the reagent probe assembly.
Patient Sequence No: 1, Text Type: N, H10
[72510616]
On 27mar2017, a customer reported an unexpected negative antibody screen when using capture-r ready indicator red cells on a galileo neo instrument, when tested on (b)(6) 2017.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1034569-2017-00070 |
MDR Report Key | 6471958 |
Date Received | 2017-04-07 |
Date of Report | 2017-04-07 |
Date of Event | 2017-03-26 |
Date Mfgr Received | 2017-03-27 |
Date Added to Maude | 2017-04-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | MEDICAL TECHNOLOGIST |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR HOWARD YOREK |
Manufacturer Street | 3130 GATEWAY DRIVE |
Manufacturer City | NORCROSS GA 300915625 |
Manufacturer Country | US |
Manufacturer Postal | 300915625 |
Manufacturer Phone | 7704412051 |
Manufacturer G1 | IMMUCOR, INC. |
Manufacturer Street | 3130 GATEWAY DRIVE |
Manufacturer City | NORCROSS GA 300915625 |
Manufacturer Country | US |
Manufacturer Postal Code | 300915625 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | CAPTURE-R READY INDICATOR RED CELLS |
Generic Name | ANTIGLOBULIN COATED RED CELLS |
Product Code | KSF |
Date Received | 2017-04-07 |
Catalog Number | 0006428 |
Lot Number | 221824 |
Device Expiration Date | 2017-04-13 |
Operator | MEDICAL TECHNOLOGIST |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | IMMUCOR, INC. |
Manufacturer Address | 3130 GATEWAY DRIVE NORCROSS GA 300915625 US 300915625 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-04-07 |