VERSACELL SAMPLE MANAGEMENT SYSTEM VERSACELL X3

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-04-07 for VERSACELL SAMPLE MANAGEMENT SYSTEM VERSACELL X3 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[72502293] The customer contacted a siemens customer care center (ccc) specialist. The ccc reviewed the data provided. The ccc found the customer's quality control (qc) was in at the time of the event. The customer stated no errors were found in the error log of the instrument. Siemens is investigating the event.
Patient Sequence No: 1, Text Type: N, H10


[72502294] Discordant sodium and chloride results were obtained on four patient samples, on an versacell x3 sample management system. The customer repeated the same samples by front loading the samples in their dimension instrument. The repeated samples resulted within the expected clinical range. The repeat results were released to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant, sodium and chloride results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2247117-2017-00047
MDR Report Key6474322
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-04-07
Date of Report2017-07-11
Date of Event2017-03-11
Date Mfgr Received2017-06-27
Date Added to Maude2017-04-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCASSANDRA KOCSIS
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242687
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer StreetREGISTRATION #: 2247117 62 FLANDERS BARTLEY ROAD
Manufacturer CityFLANDERS NJ 07836
Manufacturer CountryUS
Manufacturer Postal Code07836
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameVERSACELL SAMPLE MANAGEMENT SYSTEM
Generic NameCLINICAL CHEMISTRY GENERAL PURPOSE LABORATORY EQUIPMENT
Product CodeJGS
Date Received2017-04-07
Model NumberVERSACELL X3
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address62 FLANDERS BARTLEY RD FLANDERS NJ 07836 US 07836

Device Sequence Number: 1

Brand NameVERSACELL SAMPLE MANAGEMENT SYSTEM
Generic NameCLINICAL CHEMISTRY GENERAL PURPOSE LABORATORY EQUIPMENT
Product CodeDGC
Date Received2017-04-07
Model NumberVERSACELL X3
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address62 FLANDERS BARTLEY RD FLANDERS NJ 07836 US 07836


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.