MONTAGE SETTABLE, RESORBABLE HEMOSTATIC BONE PUTTY 900-ORT OS-MON-1001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2017-04-12 for MONTAGE SETTABLE, RESORBABLE HEMOSTATIC BONE PUTTY 900-ORT OS-MON-1001 manufactured by Abyrx Inc.

Event Text Entries

[72593135] A physician communicated that an infection had occurred at a site where montage was used. The procedure involved an incision made inside of the mouth for an orthognathic procedure on the maxilla. Montage was used on the maxilla to control bleeding in conjunction with the implantation of hardware (plate). Approximately four weeks after the surgery date the implant site had become infected, which required a surgical procedure to remove all potentially infected implant materials, treat the infection, and replace the plate. The patient was said to be doing well following this procedure. The surgeon did not remove all excess montage as instructed in the ifu and used the product in a contaminated area. The product ifu includes a precaution against placement in contaminated or infected areas. A review of the batch records for the lot did not uncover any irregularities. No other customer complaints have been reported for use of product from this lot. No information supports a product malfunction or an out of specification product. If more information becomes available it will be provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005972619-2017-00001
MDR Report Key6486872
Report SourceDISTRIBUTOR
Date Received2017-04-12
Date of Report2017-04-12
Date of Event2017-02-10
Date Mfgr Received2017-03-09
Device Manufacturer Date2016-08-30
Date Added to Maude2017-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. ANIQ DARR
Manufacturer Street1 BRIDGE STREET SUITE 121
Manufacturer CityIRVINGTON NY 10533
Manufacturer CountryUS
Manufacturer Postal10533
Manufacturer Phone9143572600
Manufacturer G1ABYRX INC
Manufacturer Street1 BRIDGE STREET SUITE 121
Manufacturer CityIRVINGTON NY 10533
Manufacturer CountryUS
Manufacturer Postal Code10533
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONTAGE SETTABLE, RESORBABLE HEMOSTATIC BONE PUTTY
Generic NameCALCIUM PHOSPHATE BONE HEMOSTASIS MATERIAL
Product CodeMTJ
Date Received2017-04-12
Model Number900-ORT
Catalog NumberOS-MON-1001
Lot Number10267
Device Expiration Date2017-08-30
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABYRX INC
Manufacturer Address1 BRIDGE STREET SUITE 121 IRVINGTON NY 10533 US 10533


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-04-12

© 2026 FDA.report