POWERED WHEELCHAIR SYSTEM INVA/MATRX MX1 BACK FITS W: 15", H: 12" IMX11512-CR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-12 for POWERED WHEELCHAIR SYSTEM INVA/MATRX MX1 BACK FITS W: 15", H: 12" IMX11512-CR manufactured by Motion Concepts Lp.

Event Text Entries

[72806518] This is the first known incident reported of our invacare-matrx mx1 back becoming disengaged from the wheelchair due to the latch hardware not engaging properly, and thereby causing the end user to fall backward from the wheelchair. The mx1 back has been on the market since 2011. At present, motion concepts is not able to complete a thorough investigation, due to a lack of information provided from the dealer. Motion concepts has requested, but to date has not been provided any images of the alleged defective hardware and/or any images to show how the mx1 back was installed on the wheelchair. The alleged defective hardware components are being returned by numotion to motion concepts for our inspection/evaluation, but have not been received at this time. Motion concepts' investigation remains open, pending receipt of additional information.
Patient Sequence No: 1, Text Type: N, H10


[72806519] On 13-march 2017, motion concepts was notified by numotion (dealer) of an alleged incident involving one of our invacare-matrx mx1 backs. Numotion was informed by the end user that her mx1 back had become disengaged suddenly from her wheelchair causing her to fall backward and bump her head. The end user did not seek medical attention. The dealer alleges the latch mechanism on the hardware was not engaging properly causing the back to become disengage. The original invacare matrx mx1 back (model: imx11512-cr) was sold to numotion in june 2016.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615350-2017-00002
MDR Report Key6487418
Date Received2017-04-12
Date of Report2017-04-12
Device Manufacturer Date2016-06-20
Date Added to Maude2017-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOCCUPATIONAL THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS DONA BHAMRA
Manufacturer Street84 CITATION DRIVE UNITS 1-7
Manufacturer CityCONCORD, ONTARIO L4K3C1
Manufacturer CountryCA
Manufacturer PostalL4K3C1
Manufacturer Phone6950134
Manufacturer G1MOTION CONCEPTS LP
Manufacturer Street84 CITATION DRIVE UNITS 1-7
Manufacturer CityCONCORD, ONTARIO L4K3C1
Manufacturer CountryCA
Manufacturer Postal CodeL4K3C1
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePOWERED WHEELCHAIR SYSTEM
Generic NameWHEELCHAIR
Product CodeKNN
Date Received2017-04-12
Model NumberINVA/MATRX MX1 BACK FITS W: 15", H: 12"
Catalog NumberIMX11512-CR
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMOTION CONCEPTS LP
Manufacturer Address84 CITATION DRIVE UNITS 1-7 CONCORD, ONTARIO L4K3C1 CA L4K3C1


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-04-12