MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-04-13 for MALYUGIN RING SYSTEM MAL-1001 manufactured by Microsurgical Technology Inc..
[72738517]
The device was returned for evaluation deformed and the evaluation was therefore unable to determine the cause of the failure.
Patient Sequence No: 1, Text Type: N, H10
[72738518]
During cataract surgery, upon insertion of an iol the haptic of the iol caught the malyugin ring scroll and pulled the ring into the capsular bag tearing it. A vitrectomy was performed and a sulcus iol was implanted. The patient was reported to be doing very well post-op.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3019924-2017-00006 |
MDR Report Key | 6490470 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2017-04-13 |
Date of Report | 2017-04-04 |
Date of Event | 2017-03-15 |
Date Mfgr Received | 2017-03-15 |
Device Manufacturer Date | 2017-01-30 |
Date Added to Maude | 2017-04-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. JOHN RAY |
Manufacturer Street | 8415 154TH AVE NE |
Manufacturer City | REDMOND WA 98052 |
Manufacturer Country | US |
Manufacturer Postal | 98052 |
Manufacturer Phone | 4255560544 |
Manufacturer G1 | MICROSURGICAL TECHNOLOGY INC |
Manufacturer Street | 8415 154TH AVE NE |
Manufacturer City | REDMOND WA 98052 |
Manufacturer Country | US |
Manufacturer Postal Code | 98052 |
Single Use | 3 |
Remedial Action | RL |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MALYUGIN RING SYSTEM |
Generic Name | CLIP, IRIS RETRACTOR |
Product Code | HOC |
Date Received | 2017-04-13 |
Returned To Mfg | 2017-03-27 |
Model Number | MAL-1001 |
Catalog Number | MAL-1001 |
Lot Number | 084930 |
Operator | PHYSICIAN |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MICROSURGICAL TECHNOLOGY INC. |
Manufacturer Address | 8415 154TH AVE NE REDMOND WA 98052 US 98052 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2017-04-13 |