DIMENSION VISTA? 1500 SYSTEM DIMENSION VISTA 1500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-04-14 for DIMENSION VISTA? 1500 SYSTEM DIMENSION VISTA 1500 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[72886483] A siemens healthcare diagnostics inc. (siemens) customer service engineer (cse) was dispatched to the customer site. The cse analyzed the instrument and replaced the s1 belt, the s2 belt, probes and a vertical motor. Verification of tbil recovery by running 400 complete metabolic panels (cmp) with repeats on any tbil over 0. 8 from the aliquot plate. All results are well within manufactures specification. The instrument is performing according to specifications. The cause of the discordant tbil results is unknown. No further action required. No further evaluation of this device is required.
Patient Sequence No: 1, Text Type: N, H10


[72886484] The customer stated that at least 70 discrepant, falsely elevated total bilirubin (tbil) results were obtained on a dimension vista 1500 instrument. The initial results were reported to the physician. The physician did not question the results. The tests were repeated on an alternate dimension vista 1500 and the results were lower. The repeated results were reported to the physician. There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated total bilirubin results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00371
MDR Report Key6494183
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-04-14
Date of Report2017-04-14
Date of Event2017-03-25
Date Mfgr Received2017-03-25
Device Manufacturer Date2008-09-24
Date Added to Maude2017-04-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARL AEBIG
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243102
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer StreetREGISTRATION #: 1226181 101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDIMENSION VISTA? 1500 SYSTEM
Generic NameDIMENSION VISTA 1500
Product CodeMQM
Date Received2017-04-14
Model NumberDIMENSION VISTA 1500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101

Device Sequence Number: 1

Brand NameDIMENSION VISTA? 1500 SYSTEM
Generic NameDIMENSION VISTA 1500
Product CodeJJE
Date Received2017-04-14
Model NumberDIMENSION VISTA 1500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-14