BLOODGAS ANALYZER ABL 725 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2005-11-30 for BLOODGAS ANALYZER ABL 725 * manufactured by Radiometer Medical Aps.

Event Text Entries

[425295] An abl725 analyzer reported a 1 point calibration drift error on the pco2 electrode. Instead of following the procedure for this kind of problem in the operator's manual (error no. 376), the user then ran four calibrations in a row which all reported the same drift problem. Then a fifth calibration was run with no drift problem, the user trusted this results, but the result made no sense to the doctor. The analyzer was then examined and a clot was found in the first chamber.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2005-00011
MDR Report Key649504
Report Source01,05,06
Date Received2005-11-30
Date of Report2005-11-28
Date of Event2005-11-05
Device Manufacturer Date2005-01-01
Date Added to Maude2005-12-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLISE-LOTTE RINGBY, SPEC.
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ 2700
Manufacturer CountryDA
Manufacturer Postal2700
Manufacturer Phone**
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ 2700
Manufacturer CountryDA
Manufacturer Postal Code2700
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLOODGAS ANALYZER
Generic NameABL 700 SERIES AND ABL 800 SERIES
Product CodeCCE
Date Received2005-11-30
Model NumberABL 725
Catalog Number*
Lot Number*
ID Number*
OperatorLAY USER/PATIENT
Device Availability*
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key638995
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* BRONSHOJ DA
Baseline Brand NameBLOODGAS ANALYZER
Baseline Generic NameABL 700 SERIES AND ABL 800 SERIES
Baseline Model NoABL 725
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-11-30
20 2005-11-30

© 2026 FDA.report