VITROS 5600 INTEGRATED SYSTEM 6802413

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-18 for VITROS 5600 INTEGRATED SYSTEM 6802413 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[73254607] The investigation determined that a lower than expected ckmb result was obtained from a patient sample processed using vitros ckmb slides on a vitros 5600 integrated system. The most likely assignable cause for the event is instrument related, as a within-run precision test was outside of ortho clinical diagnostic guidelines, indicating the vitros 5600 system was not performing as intended. The ortho field engineer performed service actions and the post service within-run precision was acceptable indicating the vitros 5600 system was performing as expected. Following service actions, acceptable vitros ckmb performance was observed.
Patient Sequence No: 1, Text Type: N, H10


[73254608] The customer observed a lower than expected ckmb result from a patient sample (2. 2 ng/ml versus expected 3. 0 ng/ml) processed using vitros ckmb reagent on a vitros 5600 integrated system. Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples. The lower than expected results were not reported from the laboratory. There were no allegations of patient harm as a result of the event. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319681-2017-00028
MDR Report Key6499527
Date Received2017-04-18
Date of Report2017-04-18
Date of Event2017-03-23
Date Mfgr Received2017-03-23
Device Manufacturer Date2013-03-14
Date Added to Maude2017-04-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal Code14626
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 0

Brand NameVITROS 5600 INTEGRATED SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeJHW
Date Received2017-04-18
Catalog Number6802413
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626

Device Sequence Number: 1

Brand NameVITROS 5600 INTEGRATED SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-04-18
Catalog Number6802413
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.