IVC BARIATRIC BED 9153641598 NA:BAR600IVC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2017-04-19 for IVC BARIATRIC BED 9153641598 NA:BAR600IVC manufactured by Invacare Florida.

Event Text Entries

[73228306] The device has been returned and an evaluation is in process. A follow-up report will be filed when the evaluation is complete. A replacement has been sent.
Patient Sequence No: 1, Text Type: N, H10


[73228307] The provider reported the end user could see the sparks coming from under the bed, they called their attendant and had the bed unplugged. When the provider went to home he noticed the plug to the pendant melted.
Patient Sequence No: 1, Text Type: D, B5


[74393773] The product was received for evaluation and the complaint of the control box sparking and emitting smoke was confirmed. The c1 capacitor shorted out causing the failure. There was evident discoloration and deformation around the d-sub 9 pin female pendant connector due to the heat source being applied from the capacitor failure. The control box housing is comprised of v0 rated material and will self-extinguish once the heat source is removed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1031452-2017-00039
MDR Report Key6503549
Report SourceDISTRIBUTOR
Date Received2017-04-19
Date of Report2017-03-21
Date of Event2017-03-20
Date Mfgr Received2017-04-27
Device Manufacturer Date2015-04-01
Date Added to Maude2017-04-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR JASON FIEST
Manufacturer StreetONE INVACARE WAY
Manufacturer CityELYRIA OH 44036
Manufacturer CountryUS
Manufacturer Postal44036
Manufacturer Phone8003336900
Manufacturer G1INVACARE FLORIDA
Manufacturer Street2101 EAST LAKE MARY BLVD
Manufacturer CitySANFORD FL 32773
Manufacturer CountryUS
Manufacturer Postal Code32773
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIVC BARIATRIC BED 9153641598
Generic NameBARIATRIC BED
Product CodeOSI
Date Received2017-04-19
Model NumberNA:BAR600IVC
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINVACARE FLORIDA
Manufacturer Address2101 EAST LAKE MARY BLVD SANFORD FL 32773 US 32773


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.