BENEVISION CENTRAL MONITORING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-19 for BENEVISION CENTRAL MONITORING SYSTEM manufactured by Shenzhen Mindray Bio-medical Electronics, Ltd..

Event Text Entries

[73216706] An on site investigation was performed by mindray service engineer and no anomaly was found on the tm80 telemetry monitoring system and the benevision cms. The log files and the patient waveform were returned for analysis and it was found that at the 22:51:49, (b)(6) 2017, there was a "ecg lead off" alarm initiated and the ecg lead was reconnected about an hour and a half later (0:18:22, (b)(6) 2017). After the ecg leads were reconnected, the cms received "asystole" alarms. At 00:24:32, (b)(6) 2017, there was an operation "press the alarm mute" in the log, which indicated the alarm sound had been turned off at that time. According to above mentioned analysis ,we concluded that the tm80 and the cms were working properly. No malfunction was found.
Patient Sequence No: 1, Text Type: N, H10


[73216707] The customer reported that during (b)(6) at 10:00 pm to (b)(6) at 2:00 am, a patient died while being monitored with the tm80 telemetry monitoring system. The benevision central monitoring system did not alarm.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007222337-2017-00011
MDR Report Key6505392
Date Received2017-04-19
Date of Report2017-03-24
Date of Event2017-03-20
Date Mfgr Received2017-03-24
Date Added to Maude2017-04-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS SERENA CHEN
Manufacturer StreetKEJI 12TH ROAD SOUTH HIGH-TECH INDUSTRIAL PARK
Manufacturer CityNANSHAN, SHENZHEN 518057
Manufacturer CountryCH
Manufacturer Postal518057
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBENEVISION CENTRAL MONITORING SYSTEM
Generic NameCENTRAL MONITORING SYSTEM
Product CodeMSX
Date Received2017-04-19
Device Availability*
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS, LTD.
Manufacturer AddressKEJI 12TH ROAD SOUTH HIGH TECH INDUSTRIAL PARK NANSHAN, SHENZHEN 518057 CH 518057


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2017-04-19

© 2026 FDA.report