MULTIDIP 8 DRUG KIT 25/PK DOA-184

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign,health profe report with the FDA on 2017-04-20 for MULTIDIP 8 DRUG KIT 25/PK DOA-184 manufactured by Alere San Diego, Inc..

Event Text Entries

[73386824] Investigation conclusion: the customer's observation was not replicated in house with retention products. Retention devices were tested with coc and thc +50% cutoff control and 3x cutoff control. All coc and thc results were positive at read time and met qc specifications. No false negative results were observed during in-house testing. Manufacturing batch record review did not uncover any abnormalities. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[73386825] The customer reported a false negative cocaine and thc result on a patient. An oral fluid sample was sent to the laboratory and produced a benzoylecgonine (cocaine) result of 63 ng/ml and a thc result of 7 ng/ml. The appropriate urine sample type was used for the multidip 8 drug test. However, oral fluid is not indicated for this test. Therefore, a direct comparison between the results provided could not be conducted. There was no treatment or intervention administered based on either set of results. No reported adverse patient sequela.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2017-00076
MDR Report Key6509463
Report SourceCONSUMER,FOREIGN,HEALTH PROFE
Date Received2017-04-20
Date of Report2017-03-24
Date Mfgr Received2017-03-24
Date Added to Maude2017-04-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMULTIDIP 8 DRUG KIT 25/PK
Generic NameENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES
Product CodeDIO
Date Received2017-04-20
Model NumberDOA-184
Lot NumberDOA6090157
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-20