ACORN 180 RH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2017-04-21 for ACORN 180 RH manufactured by Acorn Stairlifts, Inc..

Event Text Entries

[73377174] Found lift at bottom charge properly charging. Seatbelt was tangled, took seatbelt housing apart to ensure no damage to spring mechanism and untangled belt. All other components inspected to specification with exception of anti-tilt roller assembly which was broken but did not affect lift operation.
Patient Sequence No: 1, Text Type: N, H10


[73377175] Client was riding lift in upward direction and fell out when going around the curve. Client not wearing seatbelt as it was stuck. Broke her wrist and clavicle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003124453-2017-00006
MDR Report Key6511664
Report SourceCONSUMER
Date Received2017-04-21
Date of Report2017-04-21
Date of Event2017-03-13
Date Mfgr Received2017-03-23
Device Manufacturer Date2015-08-03
Date Added to Maude2017-04-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT FAMILY MEMBER OR FRIEND
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS TRACY BERO
Manufacturer Street7001 LAKE ELLENOR DRIVE
Manufacturer CityORLANDO FL 32809
Manufacturer CountryUS
Manufacturer Postal32809
Manufacturer Phone4076500216
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACORN 180 RH
Generic NamePOWERED STAIRWAY CHAIRLIFT
Product CodePCD
Date Received2017-04-21
Returned To Mfg2017-03-28
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerACORN STAIRLIFTS, INC.
Manufacturer Address7001 LAKE ELLENOR DRIVE ORLANDO FL 32809 US 32809


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2017-04-21