IDC-INTERLOCKING DETACHABLE COIL * 360202

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2005-12-05 for IDC-INTERLOCKING DETACHABLE COIL * 360202 manufactured by Boston Scientific Corp. Neurovascular Div..

Event Text Entries

[19494593] Device was advanced and pulled back into the renegade 18 microcatheter, resistance was felt. Although, attempts were made to remove coil, it got stuck at the distal tip of the microcatheter. When the coil was pulled back with force, the tip of the pusher wire got stretched and the coil was inserted into the aneurysm. No complications were reported to have occurred with the patient during this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6000078-2005-00246
MDR Report Key651342
Report Source07
Date Received2005-12-05
Date of Report2005-12-02
Date of Event2005-09-20
Date Mfgr Received2005-09-20
Device Manufacturer Date2005-06-01
Date Added to Maude2005-12-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactALEN KRIKORIAN
Manufacturer Street47900 BAYSIDE PARKWAY
Manufacturer CityFREEMONT CA 945386515
Manufacturer CountryUS
Manufacturer Postal945386515
Manufacturer Phone8006622415
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIDC-INTERLOCKING DETACHABLE COIL
Generic NameDETACHABLE COIL
Product CodeHGC
Date Received2005-12-05
Returned To Mfg2005-10-12
Model Number*
Catalog Number360202
Lot Number7737263
ID Number*
Device Expiration Date2008-06-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagY
Date Removed*
Device Sequence No1
Device Event Key640811
ManufacturerBOSTON SCIENTIFIC CORP. NEUROVASCULAR DIV.
Manufacturer Address* FREEMONT CA * US
Baseline Brand NameIDC-INTERLOCKING DETACHABLE COIL
Baseline Generic NameDETACHABLE COIL
Baseline Model No*
Baseline Catalog No360202
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2005-12-05