DIMENSION RXL MAX WITH HM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-04-22 for DIMENSION RXL MAX WITH HM manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[73515960] A siemens customer service engineer (cse) was dispatched to the customer's site. The cse performed decontamination on the instrument. The customer has not experienced any further issues. The cause of the discordant, falsely elevated total bilirubin results is unknown. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[73515961] Discordant, falsely elevated total bilirubin results were obtained on four patient samples on a dimension rxl max with hm (serial number: (b)(4)) instrument. The initial results were not reported out to the physician(s). The customer repeated the same sample on another dimension rxl max with hm (serial number: (b)(4)) instrument. The customer found a positive bias with the second instrument's results. The customer reported the initial results. There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated total bilirubin results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00423
MDR Report Key6514109
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-04-22
Date of Report2017-04-21
Date of Event2017-03-15
Date Mfgr Received2017-03-27
Date Added to Maude2017-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTIMOTHY RICE
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242406
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer StreetREGISTRATION NUMBER: 1226181 101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDIMENSION RXL MAX WITH HM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeMQM
Date Received2017-04-22
Model NumberDIMENSION RXL MAX WITH HM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 19714 US 19714

Device Sequence Number: 1

Brand NameDIMENSION RXL MAX WITH HM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-04-22
Model NumberDIMENSION RXL MAX WITH HM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 19714 US 19714


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.