MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-24 for CAPTURE-R READY INDICATOR RED CELLS 0006428 manufactured by Immucor, Inc..
[73735102]
The full number for the product in question is bk020053 (1/27/03). The product had already expired at the time of immucor's receipt of the event information. However, the immucor laboratory had previously tested retention product on 22mar2017, which had performed as expected at that time. Immucor technical support used a remote electronic connection method to assess the instrument test well images in question on (b)(6) 2017, and determined that the instrument test well image in question was visually negative, with a slight haze around the red blood cell button.
Patient Sequence No: 1, Text Type: N, H10
[73735103]
On (b)(6) 2017, an immucor employee present at a customer site reported an unexpected negative antibody screen when using capture-r ready indicator red cells on a galileo echo instrument, when tested on (b)(6) 2017.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1034569-2017-00083 |
MDR Report Key | 6515909 |
Date Received | 2017-04-24 |
Date of Report | 2017-04-24 |
Date of Event | 2017-04-06 |
Date Mfgr Received | 2017-04-10 |
Date Added to Maude | 2017-04-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | MEDICAL TECHNOLOGIST |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR HOWARD YOREK |
Manufacturer Street | 3130 GATEWAY DRIVE |
Manufacturer City | NORCROSS GA 300915625 |
Manufacturer Country | US |
Manufacturer Postal | 300915625 |
Manufacturer Phone | 7704412051 |
Manufacturer G1 | IMMUCOR, INC. |
Manufacturer Street | 3130 GATEWAY DRIVE |
Manufacturer City | NORCROSS GA 300915625 |
Manufacturer Country | US |
Manufacturer Postal Code | 300915625 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | CAPTURE-R READY INDICATOR RED CELLS |
Generic Name | ANTIGLOBULIN COATED RED CELLS |
Product Code | KSF |
Date Received | 2017-04-24 |
Catalog Number | 0006428 |
Lot Number | 221820 |
Device Expiration Date | 2017-04-06 |
Operator | MEDICAL TECHNOLOGIST |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | IMMUCOR, INC. |
Manufacturer Address | 3130 GATEWAY DRIVE NORCROSS GA 300915625 US 300915625 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-04-24 |