NSK SGS-E2S H266

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2017-04-25 for NSK SGS-E2S H266 manufactured by Nakanishi Inc..

Event Text Entries

[73624633] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[73624634] On april 6, 2017, nakanishi received an e-mail from a distributor ((b)(4)) about a handpiece overheating. Details are as follows. - on (b)(6) 2017, (b)(4) was made aware of the event by an nsk sales representative. - the event occurred on (b)(6) 2017. - a dentist was performing a surgical extraction of tooth #3 on a patient using an nsk handpiece, sgs-e2s (serial no. (b)(4)). - the patient was under local anesthesia. - during the procedure, the handpiece overheated and the patient complained about a burning sensation on the corner of the outer lip. - the dentist found a superficial epidermal burn, 6mm in diameter on the skin of the lower lip inferior lateral to commissure. - the dentist applied hawaiian moon aloe vera to the patient's skin. - no follow-up was scheduled with the patient. - the patient has recovered from the burn injury. - according to the dentist, there were no abnormalities or malfunctions observed with the handpiece prior to use.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611253-2017-00021
MDR Report Key6519384
Report SourceCOMPANY REPRESENTATIVE
Date Received2017-04-25
Date of Report2018-07-10
Date of Event2017-03-21
Date Facility Aware2017-03-27
Report Date2017-04-06
Date Reported to Mfgr2017-04-06
Date Mfgr Received2018-06-09
Device Manufacturer Date2015-06-30
Date Added to Maude2017-04-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR KENNETH BLOCK
Manufacturer Street800 E CAMPBELL RD. SUITE 202
Manufacturer CityRICHARDSON TX 75081
Manufacturer CountryUS
Manufacturer Postal75081
Manufacturer Phone9724809554
Manufacturer G1NAKANISHI INC. REGISTRATION NUMBER : 9611253
Manufacturer StreetMFR RPT# : 9611253-2017-00021 700 SHIMOHINATA
Manufacturer CityKANUMA-SHI, TOCHIGI-KEN 322-8666
Manufacturer CountryJA
Manufacturer Postal Code322-8666
Single Use3
Previous Use Code3
Removal Correction Number9611253-060818-001-R
Event Type3
Type of Report3

Device Details

Brand NameNSK
Generic NameHANDPIECE, ROTARY BONE CUTTING
Product CodeKMW
Date Received2017-04-25
Returned To Mfg2017-04-14
Model NumberSGS-E2S
Catalog NumberH266
OperatorDENTIST
Device AvailabilityR
Device Age1 YR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNAKANISHI INC.
Manufacturer Address700 SHIMOHINATA KANUMA-SHI, TOCHIGI-KEN 322-8666 JA 322-8666


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-04-25