OLYMPIC WARM SCALE 56328

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2017-04-25 for OLYMPIC WARM SCALE 56328 manufactured by Natus Medical Inc..

Event Text Entries

[73736979] The customer did not have a service history for this device, and there was no record of the unit ever being returned to natus for repair or service. Very little dust was observed in the unit, and the air inlet filter was missing. When measured, the resistance of the heating coil was within the specification resistance range of 72 +/- 2 ohms for this heating coil. The warm scale cover was then removed and the internal electronics, including fuses, pcas, wiring, battery and connections were all checked for signs of obvious damage or heat stress. Other than the heating coil, no other internal damage was observed. The unit was then powered on briefly, and it was confirmed that proper voltages were being applied to the various pcas. No overheating and/or odors were detected. The damaged/stressed heater element showed both elongation and sagging of the heater coils. The heater coil elongation resulted in the failure of the unit to regulate the temperature of the newborn warming shelf to the expected temperature of approximately 93 degrees fahrenheit. Running under steady-state conditions, the plastic warming shelf reached a temperature of 129. 6 degrees fahrenheit without the fan filter, as received from customer. When the filter was installed, the steady-state temperature of the shelf reached 141 degrees fahrenheit. It was observed that after reaching steady-state, the warm scale continuously supplied power to the heating coil. This can be explained by the fact that the heating coils were spaced much further apart in the region of the thermistor sensor assembly than in the regions where thermocouples were located, resulting in a sensed air temperature of approximately 79 degrees f, without air filter and 80. 5 degrees f with the air filter. This temperature is not high enough regulate the new born shelf temperature to 93 degrees f as suggested in the instruction manual or to trip the over-temperature cutoff value of 103 degrees. This malfunction is the result of heating coil elongation in the area of the thermistor sensor assembly. The subject device has been scrapped and given to natus qa for analysis and root cause investigation. The customer has been provided a replacement unit. Natus is currently conducting an ongoing investigation into this malfunction and will follow up with supplemental reports as additional information is obtained.
Patient Sequence No: 1, Text Type: N, H10


[73736980] On (b)(6) 2017, the customer reported that a warm scale was being returned to natus for service. The unit required calibration and had a damaged power cord. Additionally, a burning smell was detected from the unit. When the unit was received at natus for the damaged power cord and calibration, no adverse events or other malfunctions were reported. An initial determination to not report this event was determined and documented, but an investigation was launched due to the report of the burning smell. The unit was visually inspected by qa and the repair depot technician prior to the start of the investigation. Before disassembly, the warm scale was powered up for several minutes, and there was no detected burning odor. The reported power cord damage was determined to be superficial with no risk, but when the warm scale was evaluated at the natus repair depot, the returned unit showed evidence of possible thermal stress / damage to the heater element.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2017-00301
MDR Report Key6520062
Report SourceUSER FACILITY
Date Received2017-04-25
Date of Report2017-02-10
Date Mfgr Received2017-02-10
Device Manufacturer Date2006-11-29
Date Added to Maude2017-04-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJEFFREY MANTKOWSKI
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685143
Manufacturer G1NATUS MEDICAL INC.
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC WARM SCALE
Generic NameWARM SCALE 28
Product CodeFRW
Date Received2017-04-25
Returned To Mfg2017-02-17
Model Number56328
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC.
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-04-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.