POOLE WITH REMOVABLE SHEATH 0035040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2017-04-26 for POOLE WITH REMOVABLE SHEATH 0035040 manufactured by Conmed Corporation.

Event Text Entries

[74145837] Conmed corporation received two (2) "unopened" packages containing the poole with removable sheath for evaluation. The two packages were transferred to the packaging engineering lab for evaluation on 21-apr-2017. Visual examination of the returned packages found both packages open with product in the seal area of the complaint. Due to damage seen on the products, there is evidence that the devices were sealed within the sterile seal of the package. This complaint was confirmed regarding "the seal of the sterile pouch was breached. " this lot was manufactured on 03-oct-2017. A review of the manufacturing documents from the dhr/lhr has verified the devices were produced according to current and approved procedures and material specifications. Non-conformances regarding the product's identity, quality, safety, effectiveness or performance were not identified during manufacture. Of the lot containing (b)(4), there are no similar complaints received. A two (2) year review of product history for this devices included in the capa investigation showed a total of (b)(4) complaints, involving (b)(4) devices, including this complaint, regarding insufficient heat seals. During this same two (2) year time frame, over (b)(4) units were sold worldwide, making the occurrence rate for this reported failure mode (b)(4) percent. The reported packaging anomaly was obvious to the distributor and therefore prompted the return of the device for evaluation and replacement. This failure mode is addressed in the risk documents. To date, there have been no serious injuries or death related to this reported problem. Nonetheless, to prevent future recurrences, an investigation is in progress to determine the root cause and to address this reported problem. As with all medical devices, examination of packaging occurs multiple times prior to use (shipping/receiving, distribution, storage and prior to use). In addition, good clinical practice would include examination and verification of the product and its original packaging to ensure both are intact. If the product and its packaging have been opened/damaged or altered, do not use the product and contact the manufacturer immediately.
Patient Sequence No: 1, Text Type: N, H10


[74145838] The distributor in (b)(6) reported that during receiving and inspection of incoming products, two (2) packages, containing one each poole with removable sheath, were discovered with "the seal of the sterile pouch was breached. " in this instance, there was no patient involvement with this reported problem, as the packaging anomaly was discovered during inspection at the distributor facility prior to distribution to an end-user. This report is being filed based on the potential for injury with recurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1320894-2017-00102
MDR Report Key6523353
Report SourceDISTRIBUTOR,FOREIGN
Date Received2017-04-26
Date of Report2017-04-26
Date Mfgr Received2017-04-21
Device Manufacturer Date2016-10-03
Date Added to Maude2017-04-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. RACHEL SCHOTT
Manufacturer Street525 FRENCH ROAD
Manufacturer CityUTICA NY 135025994
Manufacturer CountryUS
Manufacturer Postal135025994
Manufacturer G1CONMED CORPORATION
Manufacturer Street525 FRENCH ROAD
Manufacturer CityUTICA NY 135025994
Manufacturer CountryUS
Manufacturer Postal Code135025994
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePOOLE WITH REMOVABLE SHEATH
Generic NameSUCTION INSTRUMENT
Product CodeBYY
Date Received2017-04-26
Returned To Mfg2017-04-21
Catalog Number0035040
Lot Number201610031
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCONMED CORPORATION
Manufacturer Address525 FRENCH ROAD UTICA NY 135025994 US 135025994


Patients

Patient NumberTreatmentOutcomeDate
10 2017-04-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.