DYND70800

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-04-27 for DYND70800 manufactured by Medline Industries Inc..

Event Text Entries

[74073745] Facility representative stated that a clipper smoked and was charred. Clipper was sitting in the charging base in the cath lab when staff noticed small spark and smoke coming from the charger base. A nurse unplugged the charger base and this stopped the smoking, no fire occurred. No procedures were affected due to this incident. The clipper and base have been in use for approximately three years. No staff or patient was injured and no medical intervention was required. Sample was returned and it was determined that the device was part of a corrective action that took place in 2015. It is unknown why the facility was using the recalled device. We have followed up with them to assure they do not continue to use any recalled devices. Due to the reported incident and in an abundance of caution this medwatch is being filed.
Patient Sequence No: 1, Text Type: N, H10


[74073746] It was reported that the surgical clipper charge base smoked.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1417592-2017-00026
MDR Report Key6525628
Date Received2017-04-27
Date of Report2017-07-27
Date Mfgr Received2017-04-13
Device Manufacturer Date2013-11-01
Date Added to Maude2017-04-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MEGAN DEBUS
Manufacturer StreetTHREE LAKES DRIVE
Manufacturer CityNORTHFIELD IL 60093
Manufacturer CountryUS
Manufacturer Postal60093
Manufacturer Phone8477703962
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Generic NameSURGICAL CLIPPER CHARGER BASE
Product CodeLWK
Date Received2017-04-27
Catalog NumberDYND70800
Lot Number13K1
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES INC.
Manufacturer AddressTHREE LAKES DRIVE NORTHFIELD IL 60093 US 60093


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-04-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.