MONOMAX VIOLET 2/0 (3) 70CM FR26 (M) B0041887

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-05-01 for MONOMAX VIOLET 2/0 (3) 70CM FR26 (M) B0041887 manufactured by B.braun Surgical Sa.

Event Text Entries

[74076726] (b)(4). Manufacturing site evaluation: samples received: 2 open pouches. Analysis and results: there are no previous complaints of this code batch. Manufactured and distributed in the market (b)(4) units of this code batch. There are no units in stock. These two open samples received were checked, and one of them has the aluminum pack stuck to the paper foil of second pack (the first pack sealed to second pack in the 4th sealing). This was viewed in enclosed picture with the evaluation. This defect took place in the welding machine and this unit was not sorted out by the personnel involved in this process. Final conclusion: taking into account that the one of the samples received does not fulfill the oem specifications, it is concluded that the complaint is justified. Corrective/preventive actions: this complaint will be included in the analysis of trending, and corrective action will be open if applies.
Patient Sequence No: 1, Text Type: N, H10


[74076727] Country of complaint: (b)(6). It was reported that the first pack sealed to the second pack. Which made it impossible to release the innerpack on the sterile table without help from a second person.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003639970-2017-00226
MDR Report Key6533742
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-05-01
Date of Report2017-05-01
Date Facility Aware2017-04-18
Date Mfgr Received2017-04-05
Date Added to Maude2017-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1B.BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI, BARCELONA 08191
Manufacturer CountrySP
Manufacturer Postal Code08191
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONOMAX VIOLET 2/0 (3) 70CM FR26 (M)
Generic NameSUTURES
Product CodeNWJ
Date Received2017-05-01
Returned To Mfg2017-04-11
Model NumberB0041887
Catalog NumberB0041887
Lot Number116424V027
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB.BRAUN SURGICAL SA
Manufacturer Address121 CARRETERA DE TERRASSA RUBI, BARCELONA 08191 SP 08191


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-05-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.