VIDAS? ANALYZER W2205R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-05-01 for VIDAS? ANALYZER W2205R manufactured by Biomerieux Sa.

Event Text Entries

[74167465] A customer in the us reported to biom? Rieux discrepant results associated with vidas? Analyzer (reference w2205r). The customer reported qcv failure with the vidas? Analyzer. Biom? Rieux instructed the customer to rerun the qcv check and report back with the results. The customer reported a subsequent qcv failure. Biom? Rieux requested a retrospective data analysis be completed since the last passing qcv. The customer reported back with the data analysis results. Two (2) patients were reported as negative instead of positive for mumps and four (4) patients were reported as negative instead of positive for varicella. The customer reported issuing corrected results for all six (6) patients. There is no indication or report from the laboratory or physician that the discrepant result led to any adverse event related to a patient's state of health. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00086
MDR Report Key6534921
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-05-01
Date of Report2017-05-01
Date Mfgr Received2017-04-03
Date Added to Maude2017-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CitySAINT LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameVIDAS? ANALYZER
Generic NameVIDAS? ANALYZER
Product CodeLJY
Date Received2017-05-01
Model NumberW2205R
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR

Device Sequence Number: 1

Brand NameVIDAS? ANALYZER
Generic NameVIDAS? ANALYZER
Product CodeDEW
Date Received2017-05-01
Model NumberW2205R
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.