MILLENIUM RO SYSTEM ME14714

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2017-05-03 for MILLENIUM RO SYSTEM ME14714 manufactured by Mar Cor Purification.

Event Text Entries

[74355500] The dialysis facility reported that the patient was dialyzed utilizing water that had not been processed by the portable ro system. User error is suspected as the facility accidentally connected the water hose on the dialysis machine to the disinfect port instead of the ro product water output port. Disinfectant was not attached to the hose or in the machine at the time of the incident. The patient was reported to have been on life support prior to dialysis treatment. Sometime after dialysis, the patient was removed from life support and died. The ro machine has since been quarantined, though there was no reported machine malfunction or reported correlation between the end of dialysis treatment and the patient death. The facility confirmed the feed water that was connected to the wrong port on the machine was dechlorinated and tested prior to the dialysis treatment. This complaint will continue to be monitored in mar cor purification complaint system. Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[74355501] User facility reported that a patient was being dialyzed with ro feed water. The ro system's water hose was connected to the wrong output port, thus water used in dialysis was not ro water. It has been reported to mcp the patient was on life support prior to dialysis treatment. Sometime after dialysis treatment (unknown amount of time), the patient was removed from life support and subsequently died.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019131-2017-00002
MDR Report Key6542253
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2017-05-03
Date of Report2017-05-03
Date of Event2017-04-06
Date Mfgr Received2017-04-06
Device Manufacturer Date2003-04-01
Date Added to Maude2017-05-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactALEX NELSON
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635094799
Manufacturer G1MAR COR PURIFICATION
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMILLENIUM RO SYSTEM
Generic NameWATER PURIFICATION SYSTEMS AND PORTABLE REVERSE OSMOSIS SYSTEMS
Product CodeFIP
Date Received2017-05-03
Model NumberME14714
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION
Manufacturer Address14550 28TH AVENUE NORTH MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2017-05-03