MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-05-04 for ARCHITECT PROLACTIN 07K76-25 manufactured by A.i.d.d Longford.
[74489438]
Investigation of the customer issue included a review of the complaint text, a search for similar complaints, and a review of labeling. Return material was not available. Tracking and trending did not identify an adverse trend. Labeling was reviewed and found to be adequate. Based on the available information a product deficiency of the architect prolactin reagent list number 07k76, lot number 58834ui00, was not identified.
Patient Sequence No: 1, Text Type: N, H10
[74489439]
The customer observed elevated prolactin results while using the architect prolactin assay compared to results generated using the centaur method. The customer provided the following data: architect results: 191. 27; centaur result: 26. 40. No adverse impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005094123-2017-00019 |
MDR Report Key | 6543960 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2017-05-04 |
Date of Report | 2017-05-04 |
Date of Event | 2016-05-10 |
Date Mfgr Received | 2017-04-13 |
Device Manufacturer Date | 2015-11-06 |
Date Added to Maude | 2017-05-04 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | A.I.D.D LONGFORD |
Manufacturer Street | LISNAMUCK CO. LONGFORD |
Manufacturer City | LONGFORD NA |
Manufacturer Country | EI |
Manufacturer Postal Code | NA |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARCHITECT PROLACTIN |
Generic Name | PROLACTIN |
Product Code | CFT |
Date Received | 2017-05-04 |
Catalog Number | 07K76-25 |
Lot Number | 58834UI00 |
Device Expiration Date | 2016-10-05 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | A.I.D.D LONGFORD |
Manufacturer Address | LISNAMUCK CO. LONGFORD LONGFORD NA EI NA |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-05-04 |