OPMI PICO ON S100 FLOOR STAND 000000-1095-102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-05-04 for OPMI PICO ON S100 FLOOR STAND 000000-1095-102 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[74509907] A zeiss field service engineer (fse) performed an on-site inspection and found that the screw mechanism that holds the optical head of the microscope safely to the stand was missing. The fse re-secured the optical head onto the stand with the screw mechanism and put the microscope back into service. The manufacturer reviewed the device installation process and procedures. The policies and procedures established for regular training and control of qualifications for installation and service related activities are adequate. The manufacturer concluded that the root cause of the optical head falling is improper installation performed by a non-trained person of the locally authorized distributor. The distributor was notified in writing about the unauthorized installation. The user manual (g-30-1781-en, version 7. 1, 2016-03-09, page 13) states "a zeiss service representative or an expert authorized by zeiss will install the system".
Patient Sequence No: 1, Text Type: N, H10


[74509908] The healthcare professional (hcp) reported that during examination of a patient's ear the optical head of the opmi pico on a s100 floor stand became completely loose and fell off. The microscope head made contact with the patient's head resulting in a minor bruise. There was no medical treatment required for the patient. The doctor mounted the optical head back onto the stand and completed the procedure successfully with the same microscope.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2017-00005
MDR Report Key6546625
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-05-04
Date of Report2017-05-04
Date of Event2017-04-04
Date Mfgr Received2017-04-04
Device Manufacturer Date2016-09-26
Date Added to Maude2017-05-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Remedial ActionRP
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameOPMI PICO ON S100 FLOOR STAND
Generic NameSURGICAL MICROSCOPE
Product CodeFSO
Date Received2017-05-04
Model NumberNA
Catalog Number000000-1095-102
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.