75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CML-75LN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-05-05 for 75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CML-75LN manufactured by Charter Medical, Ltd..

Event Text Entries

[74756020] The bag was retained by the user and sent to charter medical for investigation. The bag was visually inspected and a single crack was observed on the bag film located in the area below the left port tube. The film crack was located on the right side of the port and was approximately (0. 5-inch long). A film crack the size of the one observed on the returned sample would have likely resulted in a detectable leak during filling of the bag. Therefore, it can be concluded that the crack likely occurred during freezing or thawing. The root cause of the film crack could not be conclusively determined. The film crack under the left port tube suggests the bag may have encountered physical stress at some point, potentially while being handled when the bag was in the frozen state, although the exact point in the process where the film crack occurred is unknown. There are other causes that could potentially result in a film crack or tear: - overfilling the bag. An overfilled bag could create resistance during insertion and removal of the bag from the metal cassette. - excessive air left inside the bag after filling, resulting in rapid expansion during thawing. - moisture on the exterior surface of the bag or cassette interior when the bag is inserted into the metal cassette. Moisture may cause the bag film to freeze to the interior cassette surface, thus potentially resulting in damage when the frozen bag is removed from the cassette. - an unknown material anomaly. The product instruction for use sheet is provided with each cml-75ln device. Review of the ifu version provided with lot 145909 has the following precautions: - "after freezing, do not handle excessively. Port tubes and film are fragile in the frozen state and breakage may occur. Handle with care. " - "do not overfill. " - "remove as much air as possible from the container. " - "ensure bag exterior and protective freezing cassettes are dry prior to initiating freezing protocol. Moisture on the exterior of the bag or on the cassette could cause adherence of the bag to the cassette resulting in difficulty of bag removal. "
Patient Sequence No: 1, Text Type: N, H10


[74756021] The bag began leaking during water bath thawing process. Upon further inspection, the bag had developed a crack (~1 cm) near the seal where one of the ports meets the bag. There was no contact with a patient when the leak occurred. There was no injury, death or further medical intervention required as a result of this occurrence. There was no medical staff exposure as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2017-00003
MDR Report Key6549414
Date Received2017-05-05
Date of Report2017-05-05
Date of Event2017-04-05
Date Mfgr Received2017-04-05
Device Manufacturer Date2017-05-31
Date Added to Maude2017-05-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic NameCELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2017-05-05
Returned To Mfg2017-04-17
Model NumberCML-75LN
Catalog NumberCML-75LN
Lot Number145909
Device Expiration Date2019-05-01
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.