OPMI PICO ON S100 FLOOR STAND 000000-1095-102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-05-05 for OPMI PICO ON S100 FLOOR STAND 000000-1095-102 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[74656747] A zeiss field service engineer (fse) performed an on-site inspection and found that the screw mechanism that holds the optical head of the microscope safely to the stand was missing. The fse re-secured the optical head onto the stand with the screw mechanism and put the microscope back into service. The manufacturer concluded that the root cause of the optical head falling the second time is due to user error. Immediately after receiving the first report of the optical head falling on (b)(6) 2017, the doctor was advised by the local zeiss distributor not to use the system until the device is repaired by zeiss service, which was planned for and performed on (b)(6) 2017. However, the doctor decided to re-mount the optical head of the microscope onto the stand and proceeded to use the device again before it could be repaired. The user manual (g-30-1781-en, version 7. 1, 2016-03-09, page 10 and 11) states "the correct use of the device is absolutely vital for safe operation. Therefore, please thoroughly familiarize yourself with the content of these instructions for use before starting up the system". "modifications and repairs on this device or any equipment operated together with this device may only be performed by zeiss service. " the user manual (g-30-1781-en, version 7. 1, 2016-03-09, page 15, 208) further advises that "if a failure occurs which you cannot correct with the aid of the chapter 'what to do in the event of malfunctions', attach a sign to the device stating it is out of order and contact our service representative. "
Patient Sequence No: 1, Text Type: N, H10


[74656748] The healthcare professional (hcp) reported that on (b)(6) 2017, during a procedure the optical head of the opmi pico on a s100 floor stand became completely loose. The doctor was able to catch the falling head preventing it from hitting the floor or any person. The doctor mounted the optical head back onto the stand and completed the procedure successfully with the same microscope. No one was injured due to this incident. The hcp reported the same problem of the falling optical head the previous day to the local zeiss distributor and was advised not to use the system until it is repaired, which was planned for and performed on (b)(6) 2017. That incident is reported under mdr #9615010-2017-00005.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2017-00006
MDR Report Key6549515
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-05-05
Date of Report2017-05-05
Date of Event2017-04-05
Date Mfgr Received2017-04-05
Device Manufacturer Date2016-09-26
Date Added to Maude2017-05-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Remedial ActionIN
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameOPMI PICO ON S100 FLOOR STAND
Generic NameSURGICAL MICROSCOPE
Product CodeFSO
Date Received2017-05-05
Model NumberNA
Catalog Number000000-1095-102
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.