ABL 700 SERIES *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2005-09-15 for ABL 700 SERIES * manufactured by Radiometer Medical Aps.

Event Text Entries

[408506] Scalp ph of blood of fetuses can be used to decide if a cesarean section is indicated. When attempting to make blood scalp measurements for determining ph on fetuses while mother is in labor, up to 50% of samples are aborted because the blood sample is either contaminated with vaseline or silicone or because the blood sample is partly coagulated. When the measurement aborts the next sample cannot be taken immediately because of the low amount of blood available. This may cause a decision on a cesarean section to be delayed or possibly taken without the knowledge of scalp ph.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2004-00017
MDR Report Key654997
Report Source01,05,06
Date Received2005-09-15
Date of Report2004-12-10
Date Mfgr Received2003-06-23
Date Added to Maude2005-12-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLEIF GUDNITZ, QUALITY CONSULTA
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Manufacturer Phone8273228
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABL 700 SERIES
Generic NameBLOOD GAS ANALYZER WITH ADD. PARAMETERS
Product CodeCCE
Date Received2005-09-15
Model NumberABL 700
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key644480
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* * *
Baseline Brand NameABL 700 SERIES
Baseline Generic NameBLOOD GAS ANALYZER WITH ADD. PARAMETERS
Baseline Model NoABL 700
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-09-15

© 2026 FDA.report