ABL 700 SERIES ABL 725 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2005-09-15 for ABL 700 SERIES ABL 725 * manufactured by Radiometer Medical Aps.

Event Text Entries

[413228] Scalp ph of blood of fetuses can be used to decide if a cesarean section is indicated. When attempting to make blood scalp measurements for determining ph on fetuses while mother is in labor, up to 50% of samples are aborted because the blood sample is either contaminated with vaseline or silicone or because the blood sample is partly coagulated. When the measurement aborts the next sample cannot be taken immediately because of the low amount of blood available. This may cause a decision on a cesarean section to delayed or possibly taken without the knowledge of scalp ph.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2004-00016
MDR Report Key655023
Report Source01,05,06
Date Received2005-09-15
Date of Report2004-12-10
Date Mfgr Received2003-11-12
Date Added to Maude2005-12-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLEIF GUDNITZ, QUALITY CONSULTA
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Manufacturer Phone8273228
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABL 700 SERIES
Generic NameBLOOD GAS ANALYZER WITH ADD. PARAMETERS
Product CodeCCE
Date Received2005-09-15
Model NumberABL 725
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key644506
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* * *
Baseline Brand NameBLOODGAS ANALYZER
Baseline Generic NameABL 700 SERIES AND ABL 800 SERIES
Baseline Model NoABL 725
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-09-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.