VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2 1662659

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-05-08 for VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2 1662659 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[74896387] The investigation determined that lower and higher than expected vitros na+ results were obtained from quality control samples when processed on two vitros 5600 integrated systems. The most likely assignable cause for the event is user error related to improper pre-analytical calibrator fluid preparation. There was no evidence to suggest that the vitros na+ reagent or the vitros 5600 integrated systems malfunctioned.
Patient Sequence No: 1, Text Type: N, H10


[74896388] A customer obtained lower and higher than expected vitros na+ results from quality control samples when processed on two vitros 5600 integrated systems. The following results breached vitros na+ potential health and safety criteria: (b)(6). Biased results of the direction and magnitude observed could lead to inappropriate physician action if the event were to occur with patient samples. There were no unexpected patient sample results reported outside the laboratory and no allegation of patient harm due to this event. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319808-2017-00007
MDR Report Key6551355
Date Received2017-05-08
Date of Report2017-05-08
Date of Event2017-04-10
Date Mfgr Received2017-04-10
Device Manufacturer Date2016-07-18
Date Added to Maude2017-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street1000 LEE ROAD
Manufacturer CityROCHESTER NY 14606
Manufacturer CountryUS
Manufacturer Postal Code14606
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2
Generic NameIN VITRO DIAGNOSTICS
Product CodeJIX
Date Received2017-05-08
Catalog Number1662659
Lot Number0226
ID Number10758750009503
Device Expiration Date2018-07-18
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.