BACT/ALERT 3D MICROBIAL DETECTION SYSTEM 210147 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2005-12-16 for BACT/ALERT 3D MICROBIAL DETECTION SYSTEM 210147 NA manufactured by Biomerieux, Inc..

Event Text Entries

[408397] The customer called co in 2005 for assistance in clearing a string space format error on their bact/view software. When a string space format error occurs, the bact/view computer will stop communicating with a lis system until the error is cleared. The customer informed co during this call of a positive bact/alert bottle result that was not transmitted to their lis during the time before the error message was cleared. The lab was not staffed on sunday, and they had turned off their system's audible alarms. They did not realize that a bottle had gone positive, and that the bact/view computer was no longer communicating with their lis system until late morning of the same day. The customer called co back the next day in the evening to inform co that the pt who received the contaminated platelets had died. The blood bank determined that the platelets were contaminated with e-coli.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2005-00035
MDR Report Key655758
Report Source07
Date Received2005-12-16
Date of Report2005-12-13
Date of Event2005-11-20
Date Mfgr Received2005-11-21
Date Added to Maude2005-12-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactBRIAN BRUNDIDGE
Manufacturer Street100 RODOLPHE ST.
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer Phone9196202968
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameBACT/ALERT 3D MICROBIAL DETECTION SYSTEM
Generic NameMICROBIAL GROWTH MONITOR
Product CodeJTA
Date Received2005-12-16
Model Number210147
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key645246
ManufacturerBIOMERIEUX, INC.
Manufacturer Address100 RODOLPHE ST DURHAM NC 27712 US
Baseline Brand NameBACT/ALERT 3D MICROBIAL DETECTION SYSTEM
Baseline Generic NameMICROBIAL GROWTH MONITOR
Baseline Model No210147
Baseline Catalog NoNA
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2005-12-16

© 2024 FDA.report