ENSNARE ENDOVASCULAR SNARE SYSTEM EN2006020/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-05-10 for ENSNARE ENDOVASCULAR SNARE SYSTEM EN2006020/A manufactured by Merit Medical Ireland Ltd..

Event Text Entries

[74853013] The suspect device is not expected to return for evaluation. A follow up will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[74853014] The account alleges that during a biliary stent retrieval procedure the snare was damaged. When the physician removed the snare from the patient he inadvertently punctured his thumb on the damaged portion of the snare device. The damage was alleged to be at the at the distal junction weld from which all three snare loops protrude. The physician was tested for blood born pathogen exposure per hospital policy and received minor first aid for his wound.
Patient Sequence No: 1, Text Type: D, B5


[76208797] The suspect device was not returned for evaluation. The complaint could not be confirmed. The root cause could not be determined. A review of the complaint database was performed and no similar complaints were found for this lot number. A review of the device history record was performed and no exception documents were found.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616662-2017-00015
MDR Report Key6558167
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-05-10
Date of Report2017-04-19
Date of Event2017-04-19
Date Mfgr Received2017-05-19
Device Manufacturer Date2017-01-16
Date Added to Maude2017-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRADIOLOGIC TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CASEY HUGHES MS, CQE, CSQP
Manufacturer Street1600 MERIT PARKWAY
Manufacturer CitySOUTH JORDAN UT 84095
Manufacturer CountryUS
Manufacturer Postal84095
Manufacturer Phone8013164932
Manufacturer G1MERIT MEDICAL IRELAND LTD.
Manufacturer StreetPARKMORE INDUSTRIAL ESTATE
Manufacturer CityBALLYBRIT,
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENSNARE ENDOVASCULAR SNARE SYSTEM
Generic NameENDOVASCULAR SNARE SSTEM
Product CodeMMX
Date Received2017-05-10
Catalog NumberEN2006020/A
Lot NumberK1070354
Device Expiration Date2019-12-31
OperatorPHYSICIAN
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMERIT MEDICAL IRELAND LTD.
Manufacturer AddressPARKMORE INDUSTRIAL ESTATE BALLYBRIT, EI


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.