MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2017-05-10 for SILC 1116.0006S manufactured by Globus Medical, Inc..

Event Text Entries

[74889214] The bands and clamps were returned for evaluation. Upon inspection it was noted that all bands had been separated across the width and the ends had frayed. This type of band failure can only be replicated if the band has a cut or snag, and is then placed under extreme load. Evaluation of the scans provided show the clamps still in place at t8-t9 levels. It can also be noted that the screws at t10 have become loose and pulled out of the bone. It was noted that the breakage occurred in the middle of the bands, not at the clamp interface. While it is indeterminate as to the cause of the failure, it is possible the bands were nicked during insertion, frayed against hard bone, or clamp and band interference against the rod or bone. The cause of the reported issue cannot be determined.
Patient Sequence No: 1, Text Type: N, H10


[74889215] The patient complained of pain at t8-t9. The surgeon removed the silc implants and found all four bands had broken at the middle of the lamina bands. The surgeon removed the hardware.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004142400-2017-00025
MDR Report Key6558629
Report SourceCOMPANY REPRESENTATIVE
Date Received2017-05-10
Date of Report2017-04-19
Date of Event2017-04-19
Date Added to Maude2017-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL PAUL
Manufacturer Street2560 GENERAL ARMISTEAD AVENUE
Manufacturer CityAUDUBON PA 19403
Manufacturer CountryUS
Manufacturer Postal19403
Manufacturer Phone6109301800
Manufacturer G1GLOBUS MEDICAL, INC.
Manufacturer Street2560 GENERAL ARMISTEAD AVENUE
Manufacturer CityAUDUBON PA 19403
Manufacturer CountryUS
Manufacturer Postal Code19403
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSILC
Generic NameSILC BAND
Product CodeOWI
Date Received2017-05-10
Returned To Mfg2017-04-21
Model Number1116.0006S
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGLOBUS MEDICAL, INC.
Manufacturer Address2560 GENERAL ARMISTEAD AVENUE AUDUBON PA 19403 US 19403


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-05-10