MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-05-12 for DOUBLEAIRHOSE F/MINI AIR DRILL NO. 512.1 512.150 manufactured by Depuy Synthes Power Tools.
[75038448]
(b)(4). (b)(6). This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. Reliability engineering evaluated the device and the reported condition was duplicated and confirmed. The assignable root cause was determined to be due to improper maintenance. If additional information should become available, a supplemental medwatch report will be sent accordingly.
Patient Sequence No: 1, Text Type: N, H10
[75038449]
It was reported from (b)(6) that the double air hose was vibrating unusually. During service and evaluation, it was observed that the coupling tool side was defective and there was a problem with the connection air intake. It was not reported if the device was used in surgery, or if there was patient involvement. It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use. It was not reported if there were any injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 8030965-2017-12320 |
MDR Report Key | 6563330 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2017-05-12 |
Date of Report | 2017-04-24 |
Date of Event | 2017-04-19 |
Date Mfgr Received | 2017-04-24 |
Date Added to Maude | 2017-05-12 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | SERVICE AND TESTING PERSONNEL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MICHAEL COTE |
Manufacturer Street | 1302 WRIGHTS LANE EAST |
Manufacturer City | WEST CHESTER PA 19380 |
Manufacturer Country | US |
Manufacturer Postal | 19380 |
Manufacturer Phone | 6107195000 |
Manufacturer G1 | DEPUY SYNTHES POWER TOOLS |
Manufacturer Street | N/I |
Manufacturer City | N/I |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DOUBLEAIRHOSE F/MINI AIR DRILL NO. 512.1 |
Generic Name | INSTRUMENT SURGICAL, ORTHO, PNEUMATIC, POWERED AND ACCESSORY/ATTACHMENT |
Product Code | HSZ |
Date Received | 2017-05-12 |
Returned To Mfg | 2017-04-24 |
Catalog Number | 512.150 |
Lot Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | DEPUY SYNTHES POWER TOOLS |
Manufacturer Address | N/I N/I US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-05-12 |