LEVEL 1? ESOPHAGEAL STETHOSCOPE TEMPERATURE SENSOR ES400-18

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-05-12 for LEVEL 1? ESOPHAGEAL STETHOSCOPE TEMPERATURE SENSOR ES400-18 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[75045094] Customer has not returned the device to the manufacturer for device evaluation. If the device becomes available and is returned and evaluated, the manufacturer will file a follow-up report detailing the results of the evaluation.
Patient Sequence No: 1, Text Type: N, H10


[75045095] It was reported that a level 1? Esophageal stethoscope temperature sensor migrated into the patient's stomach. The patient was undergoing a laparoscopic gastric sleeve surgery at the time of the event. While the doctor was stapling the stomach (as a part of the gastric sleeve surgery), the doctor stapled over the esophageal stethoscope temperature sensor; stapling the device into the patient's stomach. To correct this, the patient's surgery had to be converted to an "open" procedure, requiring a full opening of the patient's abdomen. The event itself was determined to be an operator error. It was reported that there were no further clinical consequences to the patient.
Patient Sequence No: 1, Text Type: D, B5


[108187237]
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3012307300-2017-01047
MDR Report Key6563648
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-05-12
Date of Report2017-08-22
Date of Event2015-11-18
Date Mfgr Received2016-04-07
Date Added to Maude2017-05-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJENNIFER MENG
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone7633833078
Manufacturer G1SMITHS HEALTHCARE MFG
Manufacturer StreetS.A. DE C.V. AVE CALIDAD NO. 4 PARQUE INDUSTRIAL INTERNACIONA
Manufacturer CityTIJUANA, 22425
Manufacturer CountryMX
Manufacturer Postal Code22425
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEVEL 1? ESOPHAGEAL STETHOSCOPE TEMPERATURE SENSOR
Generic NameSTETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS
Product CodeBZT
Date Received2017-05-12
Catalog NumberES400-18
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-05-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.