[75169573] Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


[75169574] The customer reported receiving false positive cocaine and thc results for five employees. The same employees were tested multiple times and positive cocaine and thc results were produced each time. Confirmatory testing was conducted and the laboratory testing produced negative cocaine and thc results.
Patient Sequence No: 1, Text Type: D, B5


[87061222] Investigation conclusion: the customer's results were not replicated in-house with retention products of the reported lot. Retention products were tested with drug-free donor urine. All coc and thc strips produced negative results at the read time. False positive results were not observed during the investigation. Manufacturing batch record review did not uncover any abnormalities. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2017-00086
MDR Report Key6564852
Report SourceCONSUMER
Date Received2017-05-12
Date of Report2017-04-13
Date of Event2017-04-10
Date Mfgr Received2017-09-14
Date Added to Maude2017-05-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name8 PANEL INT CUP W/AD (DUE-187-071-08)
Generic NameENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES
Product CodeDIO
Date Received2017-05-12
Returned To Mfg2017-08-31
Model NumberDUE-187
Lot NumberDOA6020488
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-12