VITEK? 2 ANC TEST KIT 21347

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-05-16 for VITEK? 2 ANC TEST KIT 21347 manufactured by Biomerieux, Inc.

Event Text Entries

[75331835] A customer from the united states reported to biom? Rieux a misidentification of a clostridium septicum cap survey sample as clostridium perfrigens in association with the vitek? 2 anc test kit (udi (b)(4)). The first survey swab was subbed to initial plates, incubated for 24 hours then subbed to brucella agar and incubated for 48 hours. The isolate was re-subbed to brucella agar, incubated 48 hours and tested. A result of clostridium perfrigens (85%) was reported. After the customer received the cap results, the second swab received was subbed to initial plates, incubated for 24 hours then tested, in which the expected id of clostridium septicum (97%) was obtained. There was no patient involvement as this was a survey sample. The customer submitted lab reports from the testing of both swabs and the cap summary report for evaluation. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2017-00157
MDR Report Key6570512
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-05-16
Date of Report2017-07-11
Date Mfgr Received2017-06-13
Device Manufacturer Date2016-09-04
Date Added to Maude2017-05-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, INC
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 ANC TEST KIT
Generic NameVITEK? 2 ANC TEST CARD
Product CodeJSP
Date Received2017-05-16
Catalog Number21347
Lot Number2440100103
ID Number03573026144364
Device Expiration Date2018-03-06
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC
Manufacturer Address595 ANGLUM ROAD ST. LOUIS MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-16