COBAS 8000 C 702 MODULE C702 06473245001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-05-17 for COBAS 8000 C 702 MODULE C702 06473245001 manufactured by Roche Diagnostics.

Event Text Entries

[75322461] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[75322462] The customer stated that they had been getting low results for an unspecified number of patient samples tested for alb2 albumin gen. 2 (alb) on a cobas 8000 c 702 module (c702) between (b)(6) 2017. The customer stated that the issue has occurred multiple times when the reagent pack has less than 60 tests remaining. Quality controls will drop out of range low, then samples are repeated on a different c702 analyzer with higher results. The customer now runs controls every 4 hours due to the issue. The issue occurs only on this analyzer line, but not on their other analyzer lines. The customer stated that approximately 50 sample results were corrected. The customer provided data for two patient samples that had questionable low results. Of these two samples, one had an erroneous initial alb result that was reported outside of the laboratory. The sample initially resulted as 1. 8 g/dl. The sample was repeated on a different c702 analyzer, resulting as 2. 5 g/dl. The repeat result was believed to be correct and a corrected report was issued. The patient was not adversely affected. The alb reagent lot number was 23085601, with an expiration date of 06/30/2018. The customer states that replacing the reagent pack resolves the issue and no re-calibration is necessary. The customer was not sure if the reagent probes had been replaced since the onset of the issue. The field service engineer suspects that there is an environmental issue with the reagent compartment of the analyzer. He checked pump pressures and pipetting; no issues were identified. Alb is the only assay affected on rotor b of the analyzer. He states that values recover lower when 1/3 of the reagent is remaining. The issue is not resolved upon multiple calibration attempts and quality control runs. Controls run on a fresh uncalibrated reagent pack recover near the mean. Probe adjustments did not resolve the issue. The temperature within the reagent compartment was good. Considerable condensation was visible within rotor b. He replaced an air pump diaphragm since a hole was found. He ran precision studies, yielding excellent results. No imprecision or pipetting failure was identified. The analyzer will be further monitored.
Patient Sequence No: 1, Text Type: D, B5


[78987424] The customer stated that the issue started occurring on (b)(6) 2017. The issue occurred with both controls and patient samples only with certain reagent packs. The pack volume was typically around 60 tests when the issue occurred and remained until the pack was empty. The issue occurred 1-2 days after the pack was loaded on the analyzer. The issue was resolved once they started using a new pack. The customer tried using packs in multiple positions on the analyzer reagent rotor. The customer states that laboratory humidity and temperature were within specifications. The field service engineer also measured laboratory humidity and temperature. The customer has confirmed that the issue was resolved by the actions of the field service engineer. The customer has tried using multiple alb reagent packs and no further issues with recovery occurred. Investigations have determined that if condensed water drops into the reagent cassette, all reagents would have to be affected. The issue was resolved after replacement of the air pump diaphragm. It was concluded that preventive maintenance on the analyzer was not up to date.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2017-01019
MDR Report Key6571625
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-05-17
Date of Report2017-06-27
Date of Event2017-04-25
Date Mfgr Received2017-04-25
Date Added to Maude2017-05-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 8000 C 702 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeCJW
Date Received2017-05-17
Model NumberC702
Catalog Number06473245001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 8000 C 702 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-05-17
Model NumberC702
Catalog Number06473245001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.