API? 20 STREP 20600

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-05-17 for API? 20 STREP 20600 manufactured by Biomerieux Sa.

Event Text Entries

[75519098] A customer in france notified biom? Rieux of discrepant results associated with api? 20 strep. The customer reported that the api? 20 strep test gave a false identification. The false identification of aerococcus viridans was reported to the clinician. There was no injury to patient due to the false identification, however there was a delay greater than 24 hours because the first identification was performed on vitek? Ms and then subsequent testing on the api? 20 strep strip. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00124
MDR Report Key6572875
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-05-17
Date of Report2017-08-23
Date Mfgr Received2017-07-28
Device Manufacturer Date2015-12-14
Date Added to Maude2017-05-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CitySAINT LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX S.A.
Manufacturer Street5, RUE DES AQUEDUCS CRAPONNE, FR 69290
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPI? 20 STREP
Generic NameAPI? 20 STREP
Product CodeJSS
Date Received2017-05-17
Model Number20600
Lot Number1004525130
Device Expiration Date2016-12-13
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-17

© 2026 FDA.report