ORTHO VISION SURGICAL TABLE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-05-18 for ORTHO VISION SURGICAL TABLE manufactured by Steris Corporation - Montgomery.

Event Text Entries

[75666444] During the time of the reported event, the table top was level and the patient was secured to the table with restraints and coban tape. User facility personnel heard a "pop" sound as the patient's torso began to shift to the left side of the table. The patient was repositioned on the table and the procedure was completed successfully as no delays or cancellations were reported. A steris service technician arrived onsite to inspect the ortho vision table and found the table to be operating per specifications. The technician confirmed there was no evidence of any hydraulic leaks and did not note any issues with the operation of the table's tilt and lateral translations. The ortho vision table is not under a steris service contract and is maintained by a third-party service provider. In-service training was scheduled for 4/25/2017 however the user facility requested that steris reschedule. Steris account manager has made multiple attempts to reschedule in-service training on the proper use and maintenance of the table with the facility and the facility has not responded.
Patient Sequence No: 1, Text Type: N, H10


[75666445] The user facility reported that during a patient procedure, they heard a "popping" noise and the ortho vision table moved without being commanded to do so. The patient did not sustain any injuries.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1043572-2017-00030
MDR Report Key6576993
Date Received2017-05-18
Date of Report2017-05-18
Date of Event2017-04-19
Date Mfgr Received2017-04-19
Date Added to Maude2017-05-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CORPORATION - MONTGOMERY
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY AL 36109
Manufacturer CountryUS
Manufacturer Postal Code36109
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameORTHO VISION SURGICAL TABLE
Generic NameSURGICAL TABLE
Product CodeJEB
Date Received2017-05-18
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - MONTGOMERY
Manufacturer Address2720 GUNTER PARK DRIVE EAST MONTGOMERY AL 36109 US 36109


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-18