XCEL HIGH STRENGTH SUTURE ANCHOR (5BX) * 3910300015

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2005-12-27 for XCEL HIGH STRENGTH SUTURE ANCHOR (5BX) * 3910300015 manufactured by Stryker Endoscopy San Jose.

Event Text Entries

[424764] Metal inserter broke off into anchor. Metal rod is now inside anchor that was inserted into pt.
Patient Sequence No: 1, Text Type: D, B5


[7796719] The unit has arrived and the failure is confirmed. One metal support on the distal tip of the inserter shaft is broken off. The bend in the remaining metal support indicates high torsional force. Rotation of the shaft during implantation isnot instructed in the product ifu. The other unit returned also shows signs of torsional force. Attempting to insert the anchor into an undersized hole could cause this failure. Without details of surgery procedure and product usuage, the exact root cause cannot be determined. The assumed root cause is product misuse. This failure has never been seen in the field before and never been seen during product testing under proper usage.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2936485-2005-00080
MDR Report Key657832
Report Source05
Date Received2005-12-27
Date of Report2005-12-02
Date of Event2005-12-02
Date Mfgr Received2005-12-02
Date Added to Maude2006-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactBETH ONDERLINDE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXCEL HIGH STRENGTH SUTURE ANCHOR (5BX)
Generic NameFIXATION DEVICE
Product CodeNOV
Date Received2005-12-27
Returned To Mfg2005-12-07
Model Number*
Catalog Number3910300015
Lot Number675793
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagY
Date Removed*
Device Sequence No1
Device Event Key647305
ManufacturerSTRYKER ENDOSCOPY SAN JOSE
Manufacturer Address* SAN JOSE CA 95138 US
Baseline Brand NameXCEL HIGH STRENGTH SUTURE ANCHOR (5BX)
Baseline Generic NameFIXATION DEVICE
Baseline Model No*
Baseline Catalog No3910300015