TRULIGHT 1574851

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-05-19 for TRULIGHT 1574851 manufactured by Trumpf Medizin Systeme Gmbh + Co. Kg.

Event Text Entries

[75727918] It is known that an improper installation, maintenance or a collision can lead to the spring arm cover falling off. The device was evaluated and repaired by a trumpf service technician. The trumpf technician was informed that a third party biomed was aware of the broken elbow cover. No preventative maintenance records were found for this device.
Patient Sequence No: 1, Text Type: N, H10


[75727919] The elbow cover on an ondal spring arm, attached to the trumpf medical surgical light came off during a surgery and fell, landing in operative field. No patient injury occurred.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681407-2017-00016
MDR Report Key6579721
Date Received2017-05-19
Date of Report2017-04-26
Date of Event2017-04-26
Date Mfgr Received2017-04-26
Date Added to Maude2017-05-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFFEN ULBRICH
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer CitySAALFELD, SAALE 07318
Manufacturer CountryGM
Manufacturer Postal07318
Manufacturer Phone6715864140
Manufacturer G1TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer City07318 SAALFELD,
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTRULIGHT
Generic NameTRULIGHT 5520
Product CodeFQP
Date Received2017-05-19
Model Number1574851
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Manufacturer AddressCARL-ZEISS STRASSE 7-9 07318 SAALFELD, GM


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2017-05-19

© 2026 FDA.report