[75684326]
Patient Sequence No: 1, Text Type: N, H10


[75684327] This did not involve a patient or patient care. During preventative maintenance by the clinical engineering dept, technicians noticed a trend in a plastic bushing disintegrating. The hospital has 170 of these beds and since scrutinizing these, nine have been sequestered. Note: not all beds have been inspected- just a small sampling. One bed was repaired before this trend was noted. Technicians were not sure this would be a hazard for patients. The area where failure is occurring would be the knee/foot section of the bed. Due to the failure there is a significant amount of movement (side play) at the knee/foot section. Hill-rom representative came on site to look at the beds and since they do not have any alert/recall regarding these issues and the beds are no longer under warranty, it would be up to the hospital to repair. Technicians suggest reporting this possible/probable defect. Manufacturer response for advanta 2 hill-rom hospital bed, hill rom advanta 2 (per site reporter). Per the technician, hill rom stated there was no recall/alert regarding this issue and the beds are no longer under warranty.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6581190
MDR Report Key6581190
Date Received2017-05-22
Date of Report2017-05-12
Date of Event2017-05-05
Report Date2017-05-12
Date Reported to FDA2017-05-12
Date Reported to Mfgr2017-05-12
Date Added to Maude2017-05-22
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVANTA 2 BED
Generic NameBED, AC-POWERED ADJUSTABLE HOSPITAL
Product CodeFNJ
Date Received2017-05-22
Model NumberP1190A
Device Availability*
Device Age3 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM, INC.
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-22