JAMSHIDI BONE MARROW TRAY 11G X4 15G ASP BAK4511

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-05-22 for JAMSHIDI BONE MARROW TRAY 11G X4 15G ASP BAK4511 manufactured by Carefusion, Inc.

Event Text Entries

[75858108] (b)(4). A follow up emdr will be submitted upon carefusion's investigation. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[75858109] While doing core aspiration, (it was reported) the needle broke off in the patient's iliac crest, requiring her to go to the hospital for (surgical consult) for (possible) removal of the needle. Additional information received; hospital stated they did not see the needle after mri was performed and nothing was seen. Physician reported there was no difficulty entering the iliac crest. No further intervention needed at this point. Procedure was not repeated.
Patient Sequence No: 1, Text Type: D, B5


[78167870] (b)(4) carefusion investigation: a device history record review was completed for lot 0001062525. No issues were found, indicating that manufacturing procedures were properly followed, and subsequent quality inspections were passed. A failure could not be confirmed as no sample was available for analysis. The device history record indicated that the jamshidi bone marrow tray was built according to procedure, and quality inspections for the device were passed. Therefore, a root cause could not be identified. Based on the lack of a definitive root cause, no corrective or preventive actions were identified for the reported failure mode. This complaint will be added to the complaint trending system and will be monitored for future occurrences.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1625685-2017-00295
MDR Report Key6581296
Date Received2017-05-22
Date of Report2017-05-31
Date of Event2017-04-20
Date Mfgr Received2017-04-25
Date Added to Maude2017-05-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNA WEHRHEIM
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer G1CAREFUSION, INC
Manufacturer Street400 EAST FOSTER RD
Manufacturer CityMANNFORD OK 74044
Manufacturer CountryUS
Manufacturer Postal Code74044
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameJAMSHIDI BONE MARROW TRAY 11G X4 15G ASP
Generic NameTRAY, SURGICAL, NEEDLE
Product CodeFSH
Date Received2017-05-22
Model NumberBAK4511
Lot Number0001062525
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-22

© 2026 FDA.report