DOUBLE AIR HOSE SCHRADER STEM 3 METERS 519.51S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-05-24 for DOUBLE AIR HOSE SCHRADER STEM 3 METERS 519.51S manufactured by Depuy Synthes Power Tools.

Event Text Entries

[75824249] (b)(4). The actual device has been returned and is currently pending evaluation. Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[75824250] It was reported that during an unspecified surgical procedure right before the surgeon used the double air hose device, it was noted that the there was an air leak from the crack in the hose. According to the reporter, the device was not "function tested" during set up prior to use; therefore, they were unable to detect where the leak was coming from the hose. There were no delays to the planned surgical procedure. A spare device was used to complete the procedure successfully. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


[79166623] The actual device was returned for evaluation. Reliability engineering evaluated the device and observed that the inner hose was pulled from the plug causing it to leak. Therefore, the reported condition was confirmed. The assignable root cause was determined to be due to component damage from improper handling. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030965-2017-12539
MDR Report Key6587004
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-05-24
Date of Report2017-05-04
Date of Event2017-05-04
Date Mfgr Received2017-05-22
Date Added to Maude2017-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL COTE
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone6107195000
Manufacturer G1DEPUY SYNTHES POWER TOOLS
Manufacturer StreetN/I
Manufacturer CityN/I
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDOUBLE AIR HOSE SCHRADER STEM 3 METERS
Generic NameINSTRUMENT SURGICAL, ORTHO, PNEUMATIC, POWERED AND ACCESSORY/ATTACHMENT
Product CodeHSZ
Date Received2017-05-24
Returned To Mfg2017-05-11
Catalog Number519.51S
Lot Number0
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES POWER TOOLS
Manufacturer AddressN/I N/I US


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-24