CLINIMACS PLUS INSTRUMENT N/A 151-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2017-05-24 for CLINIMACS PLUS INSTRUMENT N/A 151-01 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[75920082] Miltenyi biotec inc. Initiated a maintenance service on 16th may 2017 which included testing the pump unit movement and tolerances. The results were laid dov/n in service report id #(b)(4). There were no deficiencies found at the time of the service. Furthermore, the cam team of miltenyi biotec inc. Also noted that they provided the customer with an extra tubing set, buffer, reagent and frozen cell sample. As a result, the customer was able to run the mock sample without any errors. Therefore, the complained failure couldn't be confirmed by miltenyi biotec.
Patient Sequence No: 1, Text Type: N, H10


[75920083] During elution step of a clinimacs? Cd34 enrichment, there was a pump stall error notification on the screen and the product was not expelled into the collection bag. The customer wanted someone to look at the process file and investigate whether there is a risk of this being a re-occurring problem. Date of event is unknown.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2017-00019
MDR Report Key6589893
Report SourceCOMPANY REPRESENTATIVE
Date Received2017-05-24
Date of Report2017-05-24
Date Mfgr Received2017-04-25
Device Manufacturer Date2016-11-11
Date Added to Maude2017-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS NANCY JOHANSEN
Manufacturer Street120 BEACON ST SUITE 200
Manufacturer CitySOMERVILLE MA 02143
Manufacturer CountryUS
Manufacturer Postal02143
Manufacturer Phone6172180062
Manufacturer G1MILTENYI BIOTEC GMBH
Manufacturer StreetFRIEDRICH-EBERT-STRASSE 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal Code51429
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINIMACS PLUS INSTRUMENT
Generic NameCLINIMACS PLUS INSTRUMENT
Product CodeOVG
Date Received2017-05-24
Model NumberN/A
Catalog Number151-01
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer AddressFRIEDRICH-EBERT-STRASSE 68 BERGISCH GLADBACH, 51429 GM 51429


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.