PRECISION MEDICAL BM7MFA1001 7MFA1001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-05-25 for PRECISION MEDICAL BM7MFA1001 7MFA1001 manufactured by Precision Medical Inc.

Event Text Entries

[75940546]
Patient Sequence No: 1, Text Type: N, H10


[75940547] The dial flowmeters in pacu without float balls are a patient safety issue. If these are set between d? Tentes there is no oxygen flow and no indication that there is no oxygen flow. Anesthesiologist noticed excessive expiratory humidity in his patient's mask and he determined that the flowmeter was sluggish and stuck between settings causing zero o2 flow. Thankfully, the patient maintained spo2 99%. Recommendation was to remove all non float ball flowmeters from service due to faulty design and replace them with flowmeters that use float balls. The dial flowmeters are being discarded. Incidentally, the dial flowmeters do have an indication written on the device that states, "no flow between settings. " this is the manufacturer's warning to users of the issue identified by the anesthesiologist.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6590229
MDR Report Key6590229
Date Received2017-05-25
Date of Report2017-05-23
Date of Event2017-02-06
Report Date2017-05-23
Date Reported to FDA2017-05-23
Date Reported to Mfgr2017-05-23
Date Added to Maude2017-05-25
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRECISION MEDICAL
Generic NameDIAL FLOWMETER, OXYGEN
Product CodeCAX
Date Received2017-05-25
Model NumberBM7MFA1001
Catalog Number7MFA1001
Device AvailabilityN
Device Age10 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPRECISION MEDICAL INC
Manufacturer Address300 HELD DR. NORTHAMPTON PA 18067 US 18067