BEACH CHAIR 800-0142

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-05-25 for BEACH CHAIR 800-0142 manufactured by Schuerch Corporation.

Event Text Entries

[75941532]
Patient Sequence No: 1, Text Type: N, H10


[75941533] Beach chair positioner was previously out for repair because of a broken back. Biomedical engineering replaced this broken part and returned positioner to the operating room (or). This beach chair positioner was placed back in service and was used for the 1st time since this repair. After positioning the patient on the o. R. Table with the beach chair positioner on the table, the nurse and surgeon noticed the replacement part was placed in the wrong position. The patient was moved off the or table to the stretcher, the staff removed the beach chair positioner, and put our 2nd beach chair positioner on the o. R. Table, and then moved the patient back to the or table. The 1st beach chair positioner was tagged and taken out of service and given to biomed to examine and fix. The biomed. Director reported his tech installed the two brackets on opposite sides and upside down. Manufacturer response for positioning aide, beach chair (per site reporter): biomed. Dept. Contacted manufacturer with a suggestion to look into the design with the goal that it not possible to have the parts appear to fit together properly when assembled. Not sure of the actual response.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6590295
MDR Report Key6590295
Date Received2017-05-25
Date of Report2017-05-12
Date of Event2017-03-06
Report Date2017-05-12
Date Reported to FDA2017-05-12
Date Reported to Mfgr2017-05-12
Date Added to Maude2017-05-25
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBEACH CHAIR
Generic NameCHAIR, SURGICAL, AC-POWERED
Product CodeGBB
Date Received2017-05-25
Model Number800-0142
OperatorPHYSICIAN
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSCHUERCH CORPORATION
Manufacturer Address452 RANDOLPH STREET ABINGTON MA 02351 US 02351


Patients

Patient NumberTreatmentOutcomeDate
10 2017-05-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.